Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual
Page 35
![background image](/manuals/600306/35/background.png)
A
Section 11
Troubleshooting Guide
Problem Cause
Solution
Optics validation
Problem with the
Repeat the procedure.
procedure shows
optical component of
If value is still out of
value outside of
the array reader
range, contact Bio-Rad
acceptable range
technical support.
Reporter validation
Problem with the
Repeat the procedure.
procedure shows
optical component of
If value is still out of
value outside of
the array reader
range, contact Bio-Rad
acceptable range
technical support.
Classify validation
Problem with the
Repeat the procedure.
procedure shows
calibration or
If values are still out of
value outside of
optical component
range, contact Bio-Rad
acceptable range
of the array reader
technical support.
Fluidics validation
Problem with
Repeat procedure. If
procedure shows
fluidics lines,
value is still out of range,
value outside
valves or sample
contact Bio-Rad
of acceptable range
needle of array reader.
technical support.
Section 12
Ordering Information
Catalog #
Description
171-203000
Bio-Plex Validation Kit 3.0, includes optics validation, fluidics
validation, reporter validation, and classify validation bead sets for
approximately 50 validation routines
171-203060
Bio-Plex Calibration Kit, includes CAL1 and CAL2 calibration beads
for approximately 50 daily calibration routines
171-203031
Bio-Plex MCV Plate II for use with Bio-Plex Manager 3.0 and
Validation kit 3.0
171-203030
Bio-Plex MCV Plate for use with Bio-Plex Manager 2.0 and
Validation kit 3.0
32
4110007c.qxd 02/25/2003 10:12 AM Page 32