Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 20

Advertising
background image

5.

Select Instrument from the main menu. Select Validation from the pull-down
menu.

6.

Enter the User name and control number. Select classify validation. Select
OK. The following dialog box appears. (Figure 8)

7.

Select the Eject button to eject the plate holder.

Fig. 10. Classify validation.

8.

Place the MCV plate II on the microplate platform.

9.

Select OK to start the classify validation procedure.

10. When the procedure is completed, values will be displayed in a dialog box as

shown below (Figure 9). The classify efficiency and DD efficiency results may
be accessed in this view.

Fig. 11. Classify validation results.

Bio-Plex Manager 3.0 and MCV Plate II Instructions

17

4110007c.qxd 02/25/2003 10:12 AM Page 17