Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 17

Advertising
background image

2

3

4

5

6

8

9

Bio

Fig. 6. Fluidics validation dialog.

4.

Add 5 drops each of fluidics beads 1 and 2 to the designated wells on the
MCV plate II.

5.

Select the Eject button in the dialog box to eject the plate holder.

6.

Place the MCV plate II in the microplate platform.

7.

Select OK to begin the fluidics validation procedure.

8.

When the procedure has been completed, results will appear in a dialog box
as shown in Figure 11:

Fig. 7. Fluidics validation results.

9.

Acceptable fluidics validation value.
% carryover

< or = 4.0%

Note: If value is not within range, repeat the procedure. If value again is not within
the specification, contact Bio-Rad technical support for assistance.

Bio-Plex Manager 3.0 and MCV Plates II Instructions

14

4110007c.qxd 02/25/2003 10:12 AM Page 14