Select Instrument from the main menu. Select Validation from the pull-down
menu. Select Optics from the pop-up menu. The following dialog box
appears:
Fig. 13. Optics validation dialog.
9.
Select the Eject button in the dialog box to eject the plate holder.
10. Place the MCV plate in the Bio-Plex microplate platform.
11. Select OK to begin the optics assessment procedure.
12. When the procedure has been completed values will be displayed in a dialog
box as shown on the next page.
13. Record the values for each optics validation parameter in the validation kit
worksheet provided with the validation kit.
Bio-Plex Manager 2.0 and MCV Plate Instructions
22
4110007c.qxd 02/25/2003 10:12 AM Page 22