Select Instrument from the main menu. Select Validation from the pull-down
menu. Select Optics from the pop-up menu. The following dialog box
appears:

Fig. 13. Optics validation dialog.

9.

Select the Eject button in the dialog box to eject the plate holder.

10. Place the MCV plate in the Bio-Plex microplate platform.

11. Select OK to begin the optics assessment procedure.

12. When the procedure has been completed values will be displayed in a dialog

box as shown on the next page.

13. Record the values for each optics validation parameter in the validation kit

worksheet provided with the validation kit.

Bio-Plex Manager 2.0 and MCV Plate Instructions

22