11. Record the median channel values in a copy of the worksheet provided with

the validation kit.

12. Repeat procedure if values are not within specifications. If any values are again

not within acceptable ranges, contact Bio-Rad technical service for assistance.

13. Acceptable reporter validation specifications:

Parameter Specification

Dynamic Range

4.15–4.28

Linearity >0.995
Slope

0.0593–0.0799

Accuracy >90%
Sensitivity

<200 MESF

10.3 Validation of Classify Efficiency

Procedure

1.

If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system. Note: Be sure to calibrate immediately before validation. Use
the High RP1 Target value for CAL2 calibration.

2.

Remove the classify validation bead set from 4°C storage and vortex each
bottle for 30 sec.

3.

Place 5 drops of each classify bead into the corresponding classify well labeled
as 34, 38, 54, 73, and 77 in the MCV plate (see Figure 2).

4.

Store stock vials at 4°C as soon as possible after use. Protect beads from
light.

5.

Select Instrument from the main menu. Select Validation from the pull-down
menu. Select Classify from the pop-up menu. The following dialog box
appears:

Fig. 17. Classify validation dialog.

6.

Select the Eject button to eject the plate holder.

Bio-Plex Manager 2.0 and MCV Plate Instructions

25

4110007c.qxd 02/25/2003 10:12 AM Page 25