Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual
Page 26
![background image](/manuals/600306/26/background.png)
Fig. 14. Optics validation results.
14. Acceptable optics validation specifications:
Parameter Specification
DD median
See insert
CL1 median
See insert
CL1 CV%
3–7%
CL2 median
See insert
CL2 CV%
4–8%
RP1 median
See insert
RP1 CV%
5–10%
If any results are not within the acceptable specifications, please contact
Bio-Rad technical service.
If any values do meet specification, repeat the procedure. If values are again
not within specifications, contact Bio-Rad technical service for assistance.
10.2 Validation of Reporter Channel Performance
Procedure
1.
If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system. Note: Be sure to calibrate immediately before validation.
Use the High RP1 Target value for CAL2 calibration.
2.
Remove the reporter validation bead set from 4°C storage and vortex each
bottle for 30 sec.
Bio-Plex Manager 2.0 and MCV Plate Instructions
23