Fig. 14. Optics validation results.

14. Acceptable optics validation specifications:

Parameter Specification

DD median

See insert

CL1 median

See insert

CL1 CV%

3–7%

CL2 median

See insert

CL2 CV%

4–8%

RP1 median

See insert

RP1 CV%

5–10%

If any results are not within the acceptable specifications, please contact
Bio-Rad technical service.

If any values do meet specification, repeat the procedure. If values are again
not within specifications, contact Bio-Rad technical service for assistance.

10.2 Validation of Reporter Channel Performance

Procedure

1.

If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system. Note: Be sure to calibrate immediately before validation.
Use the High RP1 Target value for CAL2 calibration.

2.

Remove the reporter validation bead set from 4°C storage and vortex each
bottle for 30 sec.

Bio-Plex Manager 2.0 and MCV Plate Instructions

23

4110007c.qxd 02/25/2003 10:12 AM Page 23