Protein array reader validation kit worksheet – Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 32

Advertising
background image

Bio-Plex Manager 2.0 and MCV Plate Instructions

Performed by:

I. Optical Validation

DD median

4774–6593

CL1 median

3383–4135

CL1 CV%

3–7%

CL2 median

3520–4302

CL2 CV%

4–8%

RP1 median

15205–18583

RP1 CV%

5–10%

II. Reporter Validation

B(Blank)

1

2

3

4

5

III. Classify Validation

Region 34

>80%

Region 38

>80%

Region 54

>80%

Region 73

>80%

Region 77

>80%

IV. Validation of Fluidics Integrity

Fluidics Integrity Validated? (Place “X” in appropriate box)

To Generate Report:

1.

Open Excel and open the ValidationReport.xls file.

2.

Click on 'Worksheet' tab and enter values from this worksheet.

3.

Click on Report and select 'File' then 'Print' to print report.

Protein Array Reader

Validation Kit Worksheet

4106178 Rev A

Parameter

Specification

Measured Value

Region

Specification

Measured Value

YES

NO

Bead

Median Channel

29

4110007c.qxd 02/25/2003 10:12 AM Page 29