Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual
Page 30
2.
Select the Unclog icon from the toolbar.
3.
The following dialog box appears:
Fig. 19. Unclog dialog.
4.
Add 5 drops of CAL1 beads to the CAL1 well on the MCV plate.
5.
Fill the DI H
2
O and 70% isopropanol reservoirs.
6.
Select the Eject button in the dialog box to eject the plate holder.
7.
Place the MCV plate in the Bio-Plex microplate platform.
8.
Select OK to begin the validation of fluidics integrity procedure.
9.
If the unclog procedure was successful, the following message will appear:
10. Check "yes" in Section IV of the validation kit worksheet.
11. If the procedure is unsuccessful, the following message will appear: If this is
the case, repeat steps 4–11.
12. If the procedure is again unsuccessful, contact Bio-Rad technical support for
assistance.
Bio-Plex Manager 2.0 and MCV Plate Instructions
27
4110007c.qxd 02/25/2003 10:12 AM Page 27