Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 30

Advertising
background image

2.

Select the Unclog icon from the toolbar.

3.

The following dialog box appears:

Fig. 19. Unclog dialog.

4.

Add 5 drops of CAL1 beads to the CAL1 well on the MCV plate.

5.

Fill the DI H

2

O and 70% isopropanol reservoirs.

6.

Select the Eject button in the dialog box to eject the plate holder.

7.

Place the MCV plate in the Bio-Plex microplate platform.

8.

Select OK to begin the validation of fluidics integrity procedure.

9.

If the unclog procedure was successful, the following message will appear:

10. Check "yes" in Section IV of the validation kit worksheet.

11. If the procedure is unsuccessful, the following message will appear: If this is

the case, repeat steps 4–11.

12. If the procedure is again unsuccessful, contact Bio-Rad technical support for

assistance.

Bio-Plex Manager 2.0 and MCV Plate Instructions

27