Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 29

Advertising
background image

2

3

4

5

6

8

9

Bio

7.

Place the MCV plate on the Bio-Plex microplate platform.

8.

Select OK to start the classify validation procedure.

9.

When the procedure is completed, values will be displayed in a dialog box as
shown below:

Fig. 18. Classify Validation dialog.

10. Record the classify efficiency values in a copy of the validation kit worksheet

provided in this manual.

11. Acceptable classify efficiency values:

Classify Bead

Specification

34 >80.0%
38 >80.0%
54 >80.0%
73 >80.0%
77 >80.0%

Note: If any values do not meet specifications, repeat the procedure. If
values are again not within specifications, contact Bio-Rad technical service
for assistance.

10.4 Validation of Fluidics Integrity

Procedure

1.

If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system. Note: Be sure to calibrate immediately before validation.
Use the High RP1 Target value for CAL2 calibration.

2.

Remove the calibration bead set from 4°C storage and vortex each bottle for
30 sec.

Bio-Plex Manager 2.0 and MCV Plate Instructions

26

4110007c.qxd 02/25/2003 10:12 AM Page 26