Fig. 5. Optics validation results.
14. Acceptable optics validation specifications:
Parameter Specification
DD median
See insert
CL1 median
See insert
CL1 CV%
3–7%
CL2 median
See insert
CL2 CV%
4–8%
RP1 median
See insert
RP1 CV%
5–10%
If any values do meet specification, repeat the procedure. If values are again
not within specifications, contact Bio-Rad technical support for assistance.
9.3 Validation of Fluidics Integrity
Procedure
1.
If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system.
2.
Select instrument, then validation from the main menu and select fluidics in
the dialog box.
3.
Enter user name and control number. Select OK, the following dialog box
appears (Figure 10).
Bio-Plex Manager 3.0 and MCV Plate II Instructions
13
4110007c.qxd 02/25/2003 10:12 AM Page 13