Bio-Rad Bio-Plex System Validation and Calibration Tools User Manual

Page 16

Advertising
background image

Fig. 5. Optics validation results.

14. Acceptable optics validation specifications:

Parameter Specification

DD median

See insert

CL1 median

See insert

CL1 CV%

3–7%

CL2 median

See insert

CL2 CV%

4–8%

RP1 median

See insert

RP1 CV%

5–10%

If any values do meet specification, repeat the procedure. If values are again
not within specifications, contact Bio-Rad technical support for assistance.

9.3 Validation of Fluidics Integrity

Procedure

1.

If not already done, follow the procedure for start-up and calibration of the
Bio-Plex system.

2.

Select instrument, then validation from the main menu and select fluidics in
the dialog box.

3.

Enter user name and control number. Select OK, the following dialog box
appears (Figure 10).