Dako Artisan™ Link

|

Basic User Guide

39

To prime the reagent packs:

1. Remove the dispense tip cover.
2. Press the dispenser tip button while holding the reagent pack over a secondary waste

container (e.g., a plastic container).

3. Press until a full stream of reagent flows from the pack. This removes any air that may be

trapped in the dispenser tip.

Load Reagents onto the Reagent Carousel

Now you are ready to load the reagents onto the Artisan™ Link

1. Using the printed

Required Reagent Report as a guide, place the required reagent packs onto

the reagent carousel.

Note:

The instrument starts scanning reagents at position 1. Therefore, you should start

loading reagents at position 1.

2. When reagent scanning begins, the instrument scans the reagent positions in the carousel

starting with position 1. It will continue scanning from that point until it encounters three (3)

empty reagent positions. It then stops scanning for reagents.

DakoLink Bar Code Scanner

You can check the Reagent pack history without having to place the pack onto the reagent carousel

by using the

DakoLink Bar Code Scanner.

Hold the scanner 1-4 inches from the small square DataMatrix bar code on the reagent pack

label.

Squeeze the trigger on the scanner until you hear a beep.

The bottle detail screen will appear.