Monthly routine maintenance, Clean the bulk fluid bottles, Flush bulk liquid lines with alcohol – Dako Artisan™ Link Pro User Manual

Page 71: Annual preventative maintenance

Advertising
background image

Dako Artisan™ Link

|

Basic User Guide

63

CAUTION

Frequent accumulation of liquids in the spill tray indicates a leak or

aspiration problem with the instrument. If you observe such accumulation, contact

Dako’s Technical Service Group.

1. Carefully pull the spill tray out from the bottom of the instrument for inspection.
2. Completely remove the tray.
3. If any liquid has collected in the spill tray, use dry gauze to absorb it.
4. Wipe with gauze dampened with deionized water or 1X Wash Solution to clean the tray. You

may use a 10% bleach solution if the plastic has been stained by dye.

Monthly Routine Maintenance

Monthly Routine Maintenance includes cleaning bulk fluid bottles.

Clean the Bulk Fluid Bottles

Keep all

bulk fluid bottles clean to perform properly. Dako Wash Solution does not contain

preservatives. If the cleanliness of these bottles is not maintained, contaminants may grow in the

lines and affect the quality of the staining results. All bulk fluid bottles should be washed according

to good laboratory practices each time before they are filled with fresh reagent. Dako recommends

rinsing them with alcohol to remove any potential contaminants.

Note:

Do not use bleach in or on Artisan™ Link,

except for spill tray. Bleach may

react with other chemicals and create toxic fumes.

Flush Bulk Liquid Lines with Alcohol

Materials needed: bulk liquid bottle with 95% Ethanol (70% Ethanol optional)

Flushing the bulk liquid line clears accumulated deposits from the lines and minimizes

contamination.

1. Connect the bulk liquid bottle filled with 95% Ethanol (70% Ethanol optional) to assigned

position.

2. Prime assigned position a minimum of three times.
3. Repeat steps 1-2 for remaining bottle positions by reusing the bottle at the next bottle

position.

4. Refill the remaining cleaned bottles with the appropriate bulk liquid and replace them on the

Artisan™ Link.

5. Click the

Prime All button.

6. Repeat step 5 five times.

Annual Preventative Maintenance

Annual maintenance must be performed by a trained Dako technician. Contact your local Dako

representative for further information.