Vital Signs Monitor

23

When the desired alarm limit is displayed, pressing SELECT ALARM again saves the displayed
alarm limit and changes the alarm parameter. Alternatively, pressing no buttons for approximately
10 seconds will cause the displayed value to be set as the alarm limit, and the Monitor will automat-
ically revert to the current operating mode.

Blood Pressure - Low Systolic Limit

Press the SELECT ALARM button until the SYSTOLIC display is illuminated, and the CYCLE display
shows "LO." This indicates the low systolic alarm setting.

The systolic display will show the current alarm limit for low systolic blood pressure in mmHg or
three dashes ("- - -") if the alarm is turned off. Pressing the SET (arrow up and arrow down) buttons
will cycle through the options in 5 mmHg intervals. The operator may choose the low systolic blood
pressure limit within the range of the current high systolic alarm limit through 65 mmHg. The alarm
may be turned off by choosing "- - -."

When the desired alarm limit is displayed, pressing SELECT ALARM again saves the displayed
alarm limit and changes the alarm parameter. Alternatively, pressing no buttons for approximately
10 seconds will cause the displayed value to be set as the alarm limit, and the Monitor will automat-
ically revert to the current operating mode.

Blood Pressure - High Diastolic Limit

Press the SELECT ALARM button until the DIASTOLIC display is illuminated, and the CYCLE display
shows "HI." This indicates the high diastolic alarm setting.

The diastolic display will show the current alarm limit for high diastolic blood pressure in mmHg or
three dashes ("- - -") if the alarm is turned off. Pressing the SET (arrow up and arrow down) buttons
will cycle through the options in 5 mmHg intervals. The operator may choose the high diastolic
blood pressure limit within the range of the current low diastolic alarm limit through 155 mmHg.
The alarm may be turned off by choosing "- - -."

When the desired alarm limit is displayed, pressing SELECT ALARM again saves the displayed
alarm limit and changes the alarm parameter. Alternatively, pressing no buttons for approximately
10 seconds will cause the displayed value to be set as the alarm limit, and the Monitor will automat-
ically revert to the current operating mode.

Blood Pressure - Low Diastolic Limit

Press the SELECT ALARM button until the DIASTOLIC display is illuminated, and the CYCLE display
shows "LO." This indicates the low diastolic alarm setting.

The diastolic display will show the current alarm limit for low diastolic blood pressure in mmHg or
three dashes ("- - -") if the alarm is turned off. Pressing the SET (arrow up and arrow down) buttons
will cycle through the options in 5 mmHg intervals. The operator may choose the low diastolic
blood pressure limit within the range of 35mmHg through the current high diastolic alarm limit. The
alarm may be turned off by choosing "- - -."

When the desired alarm limit is displayed, pressing SELECT ALARM again saves the displayed
alarm limit and changes the alarm parameter. Alternatively, pressing no buttons for approximately
10 seconds will cause the displayed value to be set as the alarm limit, and the Monitor will automat-
ically revert to the current operating mode.