Welch Allyn Vital Signs Monitor - User Manual User Manual

Page 52

Advertising
background image

46

Vital Signs Monitor

7 MONITOR WILL NOT

TURN ON

Low battery

Check connections between monitor and trans-
former, and transformer and wall receptacle.

Monitor not powering up

Unplug unit from wall receptacle and check for
breaks in cord. If connections secure, check
electrical outlet

Charging light will be on if connections are OK
and the monitor is plugged into a working out-
let

Notify Biomedical Department or Welch Allyn
Technical Support.

8 CUFF TOO TIGHT

(Over inflation)

Pressure Preset too high

Check default Pressure Preset setting: (in Con-
figuration Mode)

Unless patient has underlying systolic hyper-
tension, set pressure preset at 160 mmHg. (If
systolic blood pressure greater than pressure
preset, monitor will automatically increase an
additional 40 mmHg)

9 CUFF POPPING OFF

Inappropriate size cuff

Determine cuff size with the cuff markings or
refer to chart in Operator’s Manual for deter-
mining cuff size.

If cuff continues to pop off, notify Biomedical
Department or Welch Allyn Technical Support.

Cuff applied inside out

Re- apply cuff:

Make sure Welch Allyn label is facing away from
arm.

10 CUFF DEFLATING

TOO SLOWLY

Normal operation

Typical time to take a reading is 20 to 45 seconds.
165 seconds is the maximum.

Pressure Preset too high

Check default Pressure Preset setting. (in Con-
figuration Mode)

Patient movement

Have patient sit still.

Do not have arm tight against chest wall as res-
pirations may affect speed and accuracy of
blood pressure measurement.

Small leak in pneumatic
system

Check cuff and tubing for leaks.

Symptom

Possible Cause

Explanations and Corrective Action