Vital Signs Monitor

45

2 CUFF INFLATION &

DEFLATION WITH
NO BLOOD PRES-
SURE READING DIS-
PLAYED

(or error code in display)

Leak in pneumatic system

Ensure all cuff attachments are tight.

Carefully check for tubing leaks in blood pres-
sure cuff and tubing attachment to monitor.

Arm movement during
cycle

Keep arm still during blood pressure cycle:

movement may cause inaccuracies from arti-
fact

Tubing movement artifact

Do not contact tubing during blood pressure
cycle:

movement may cause inaccuracies from arti-
fact

3 NO CUFF INFLATION

Connections from monitor
to cuff loose

Check all connections. (Do not over tighten).

4 TEMPERATURE

MALFUNCTION

Broken probe

Replace probe.

Consult Technical Manual.

Notify Biomedical department or Welch Allyn
Technical Support.

Wait for display window to read OrL before
placing probe.

Improper probe placement

Place probe in most posterior sublingual
pocket.

Notify Biomedical department or Welch Allyn
Technical Support.

Probe not replaced

Replace probe in holder prior to taking another
temperature.

5 SpO

2

MALFUNC-

TION

Sensor in Place but No
SpO

2

on Display

Inaccurate SpO

2

Reading

Improperly attached sensor

Insert the patient’s finger completely into sen-
sor

Cable incorrectly plugged
into monitor

Ensure sensor cable is correctly plugged into
monitor

Sp0

2

disabled

Ensure Sp02 is enabled (check Configuration
Mode)

Incorrect Sensor

Ensure that correct manufacturers sensor is in
use

Nellcor sensors are not interchangeable

6 PRINTER MALFUNC-

TION

Paper will not advance

Consult Technical Manual

Notify Biomedical department or Welch Allyn
Technical Support.

Symptom

Possible Cause

Explanations and Corrective Action