Blood pressure automatic mode, Temperature operating modes – Welch Allyn Vital Signs Monitor - User Manual User Manual

Page 13

Advertising
background image

Vital Signs Monitor

7

A measurement cycle may be cancelled at any time by pressing the CANCEL button. This action
immediately initiates a rapid cuff deflation.

The blood pressure measurement data will appear on the display immediately following the mea-
surement. The display will blank after two minutes. If the display is blanked, pressing the REVIEW
button on the front panel will recall the measurement. Up to 99 prior measurements are available
for review or printing.

In Blood Pressure Manual Mode, the following alarm limits may be activated; SYSTOLIC HIGH
LIMIT, SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW LIMIT, PULSE RATE
HIGH LIMIT and PULSE RATE LOW LIMIT. Blood pressure determinations which activate alarms
are indicated by flashing displays and a repetitive audible tone. If an alarm limit is violated, subse-
quent blood presssure determinations may be made only after the alarm condition is reset by press-
ing any button on the Monitor's display.

Blood Pressure Automatic Mode

The Automatic Blood Pressure Mode is entered by pressing the AUTO button. Pressing the AUTO
button displays a choice of 11 cycle interval times as follows: "St" (STAT mode), 1 minute, 3 min-
utes, 4 minutes, 5 minutes, 10 minutes, 15 minutes, 30 minutes, 45 minutes, 60 minutes and 90
minutes. These choices represent the time interval from the beginning of one cycle to the begin-
ning of the next automatic cycle. STAT mode allows the monitor to take continuous blood pressure
measurements for 15 minutes. In addition, the operator may choose to disable the Auto Mode by
choosing "--" (two dashes).

NOTE:

In the STAT and 1-minute automatic cycle mode, the Welch Allyn Vital Signs Monitor
will automatically take blood pressure measurements in 1-minute intervals for up to 15
minutes.

A measurement cycle may be cancelled at any time by pressing the CANCEL button. This action
immediately initiates a rapid cuff deflation.

The blood pressure measurement data will appear on the display immediately following the mea-
surement and will remain displayed until the next measurement cycle is initiated.

In Blood Pressure Auto Mode, the following alarm limits may be activated; SYSTOLIC HIGH LIMIT,
SYSTOLIC LOW LIMIT, DIASTOLIC HIGH LIMIT, DIASTOLIC LOW LIMIT, PULSE RATE HIGH
LIMIT and PULSE RATE LOW LIMIT. Blood pressure determinations which activate alarms are indi-
cated by flashing displays and a repetitive audible tone. Any alarm limit violation must be reset to
continue automatically timed blood pressure determinations. The alarm may be reset by pressing
any button on the Monitor's display.

Temperature Operating Modes

Thermometry measurements are made with the integrated Welch Allyn SureTemp® thermometer.
Oral and rectal probes utilize single-use disposable probe covers which limit cross-contamination.
Oral, axillary or rectal temperatures are taken using 'Normal' or 'Monitor' operating modes. Axillary
temperatures are taken using the oral probe.

In Normal mode the thermometer's microprocessor "predicts" body temperature in about 4 seconds
for oral temperatures, about 10 seconds for axillary temperatures and in about 15 seconds for rectal