Vital Signs Monitor

69

Color
Oral/Axillary Temperature Probe - Blue
Rectal Temperature Probe - Red

Mounting
Self-Supporting on rubber feet
IV Pole Mountable
Custom Mobile Stand
Wall Mountable
Attaches to Bed Rail

Portability

May be hand carried when held by the recessed handle.

When attached to an IV pole, or mounted on its custom mobile stand, the monitor and accessories
can be wheeled from patient to patient.

When attached to the bed rail may be transported with the patient within the hospital environment.

Electrical Specifications

Power Requirements
Patient-Rated isolation transformer is connected to AC mains:

North American Version:

120VAC, 60Hz. 0.20A Input, 8VDC, 0.75A Output

International Version:

240VAC, 50Hz 0.10A Input, 8VDC, 0.75A Output

Australian Version:

240VAC, 50Hz, 13VA Input, 8VDC, 0.75A Output

Battery
Lead acid, with external recharge capability.

Battery will fully charge to at least 90% capacity in 12 hours. Unit will operate and charge battery
simultaneously when connected to power source.

A fully charged battery will support 200 "typical" blood pressure determinations taken at 3 minute
intervals. Battery is 90-100% charged after 12 hours of charging. The battery automatically charges
when the monitor is powered through the AC power transformer. The battery will charge faster
when the instrument is not in operation.

Environmental Specifications

Operating Temperature
+10°C to +40°C
+50°F to +104°F
*Exception: Thermometry module will not maintain its performance characteristics below 60°F

(16°C).

Operating Altitude
-170m to +4877m
-557ft. to +16,000ft.