4. Attach the battery connectors to the new battery as shown below.

5. Slide the new battery into the battery compartment as far as it will go.

6. Replace the battery door, tightening each of the (4) screws.

7. Connect the AC power transformer to the monitor and allow the new battery to charge for

approximately 16 hours. The Monitor may be used during this charging period. While charging,
the charging indicator will flash a yellow light. When the Monitor is fully charged, the flashing
light will be replaced with a solid yellow light.

8. The battery is a non-spillable lead-acid battery. In the USA, call 1-800-SAV-LEAD for instructions

on how to recycle. For International users, contact your local authorities on recycling.