13

C. CANCEL Button

If a blood pressure cycle is in progress, pressing this button aborts the cycle and immediately
releases the cuff pressure. In "STAT" automatic blood pressure mode, pressing the cancel button
aborts the cycle in process, and cancels the "STAT" mode as well.

When pressed for 3 seconds, the cancel button erases all stored data in the Welch Allyn Vital
Signs Monitor.

D. REVIEW Button

Pressing this button will display the most current set of stored data including BP & pulse rate
and/or SpO

2

and/or temperature. Subsequent presses will display the next most recent data

sets.

E. AUTO Button & Indicator

Pressing this button allows the operator to scroll through the automatic blood pressure interval
options. Time intervals are displayed in minutes on the CYCLE display (see below). Choosing
any time interval except "--" will cause the Auto indicator light to glow, and blood pressure deter-
minations to be made according to the displayed time interval.

F. SILENCE Button & Indicator

During SpO

2

monitoring, should an SpO

2

or Pulse Rate alarm condition occur, pressing the

Silence button invokes 'Silence Mode'. This will silence the audible tone (display will continue to
flash), while the practitioner attends to the patient and Monitor. Silence mode resets automati-
cally after 30 seconds, or when the patient's condition returns to within the preset alarm param-
eters. In Silence Mode, the silence indicator light will glow.

G. SELECT ALARM Button

Pressing this button allows the operator to select a specific parameter for which an alarm thresh-
old will be set. The HIGH ("HI") or LOW ("LO") indicator will show in the CYCLE display, and the
current value will appear in the respective display.

H. SET Buttons (Arrow Up or Arrow Down)

These buttons increment or decrement the currently displayed alarm threshold.

The SET-Arrow Up button has additional functionality. See“Using the RS232 Computer Inter-
face” on page 40

I.

PULSE TONE Indicator
This indicator will be lit when the pulse tone function is enabled.

J. PULSE TONE Button

Pressing this button allows the operator to control the volume of the SpO

2

pulse tone.

K. SYSTOLIC Display

This red LED display shows the systolic blood pressure.

Additionally, this display can show systolic alarm limits (see above).