Troubleshooting

Error Indications and Interpretation

The following table of alarm conditions and error codes is intended to provide the operator with a
quick reference to the descriptions and probable causes of the error codes. For service level trou-
bleshooting, refer to the service manual.

When responding to a Monitor alarm, always check the PATIENT FIRST and then proceed to check
the Monitor.

Press CANCEL to reset patient alarm conditions.

ERROR CODE

MEASUREMENT MODE

DESCRIPTION

CORRECTIVE ACTION

C01

Blood Pressure

Measurement cycle was
cancelled by operator

-------

C02

Blood Pressure

Autozero Failure

Check for air obstruction
limit patient movement

C03

Blood Pressure

Inflation too rapid

Check for kinked hose or
air obstruction

C04

Blood Pressure

Excessive inflation time

Check for air leaks

C05

Blood Pressure

Excessive noise

Check patient condition,
cuff placement, limit
patient movement

C06

Blood Pressure

Measurement was out-
side of Monitor's measure-
ment range

Check patient condition

E10

Blood Pressure

Cuff overpressure condi-
tion

Check patient condition

C20

Temperature

Broken/missing probe

Replace probe

C21

Temperature

Probe not ready

Replace probe in holder,
wait 15 seconds, retry

E0.0 thru E9.9

Temperature

Internal malfunction

Contact customer service

“P”

Temperature

Loss of tissue contact

Ensure proper probe posi-
tioning

“--”

SpO

2

Sensor Error

Check patient condition,
sensor position and con-
nection

E7

SpO

2

Internal Malfunction

Contact customer service

C8

SpO

2

Bad sensor

Replace sensor

E11

General

Internal safety violation

Adjust temperature or
monitor location