Welch Allyn Vital Signs Monitor - User Manual User Manual

Page 73

Advertising
background image

Vital Signs Monitor

67

Overpressure Cutoff
295 - 330 mmHg

Temperature Range
84.0 - 108.0°F (28.9 - 42.2°C)

Temperature Calibration Accuracy

Calibration Accuracy: ± 0.2° F (± 0.1° C)

Temperature Determination Time
Normal Mode: ORAL: 4 seconds typical, Axillary: 10 seconds typical
Rectal: 15 seconds typical
Monitor Mode: ORAL/Rectal: 3 minutes, Axillary: 5 Minutes

Oxygen Saturation Range (SpO

2

%)

40 - 99% oxygen saturation

SpO

2

Accuracy

Without Motion - Adults: 70 - 99% ± 2 digits*
With Motion - Adults: 70 - 99% ± 3 digits**
Low Perfusion: 70 to 99% ± 2 digits***
<70% unspecified by the OEM

Biocompatibility testing has been conducted on Nellcor sensors in compliance with ISO10993-1,
Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing. The sensors have passed
the recommended biocompatibility testing and are therefore in compliance with ISO 10993-1.

*

Adult specifications are shown for OxiMax MAX-A sensors. Saturation accuracy will vary by
sensor type. Refer to the following Sensor Accuracy Grid.

**

Applicability: OxiMax MAX-A, MAX-P, and MAX-I sensors.

***

Specification applies to monitor performance and was validated with Biotek and Nellcor
simulators.