CAUTION!
The conversion of the "ANALYSETTE 3 PRO" to the micro

mill "PULVERISETTE 0" cannot be recommended, as the

amplitude control is unusable due to the ball collision mal-

functions and therefore has no utility value. In this case, the

"ANALYSETTE 3" SPARTAN is more suitable.
However if you are using the PRO version with the PULVERI-

SETTE 0 grinding attachment, note that the amplitude must

not be set to > 2 mm, as otherwise damage will be caused

to the device, for which we accept no responsibility.

The "ANALYSETTE 3 SPARTAN" is equivalent in design to the PRO model,

except that the range of functions is limited. The serial interface, elec-

tronic amplitude control, amplitude display, interval time specification,

micro sieving option as well as the standby function are not included.

2.2.1 Operating principle

The plastic housing contains a solid cast structure with an electronically

controlled electromagnet. Three permanently flexible flat springs carry

the pole plate of this magnet together with the vibratory plate, which is

fastened to it. When the magnet is switched on, the pole plate and vibra-

tory plate are attracted and spring back when it is switched off. The cast

structure and magnet on one side and the pole plate, vibratory plate and

sieves on the other side form a vibratory system.
The different number of sieves and sieving stocks or grinding attach-

ments changes the natural frequency of the vibratory system. For this

reason, it is not always possible to set the amplitude of the sieve shaker

optimally e.g. with a constant 50 Hz mains frequency present. The pro-

cessor-controlled electronics of the ANALYSETTE 3 PRO sieve shaker

ensures a precisely adjustable, reproducible sieve amplitude. This is ach-

ieved by bringing the frequency to stimulate oscillation closer to the nat-

ural frequency of the system or farther away from it. The desired oscilla-

tion amplitude of the sieve stack between 0.1 mm and 3.0 mm can

always be achieved - and with relatively low energy input. Therefore,

continuous operation is possible without heating the sample material

and the overall sieve system.

2.2.2 Amplitude control

An electronic control circuit guides the sieve frequency of the ANALY-

SETTE 3 PRO from a high to a low frequency range. In the meantime, a

measuring system records the amplitude and reports it to the control cir-

cuit until the preselected amplitude is reached.
This amplitude control takes place at regular intervals during the entire

operation. This makes it possible to react to changes in the sieve system.