6.1.3 Feed quantity for dry or wet sieving

Maximum load:
‐ Sieving stock < 1 kg
‐ Sieves and sieving stock < 6 kg

6.2 Dry sieving

For dry sieving, use the dry sieve pan (without outlet) and the dry

clamping lid (plexiglas clamping lid without nozzles).

6.2.1 Sieving parameters

Parameters

Coarse sample

Fine sample

Sieving time

3...20 min

15....30 min

Amplitude

2.5....3 mm

1.5....2.5 mm

In order to achieve an amplitude set point of 3 mm (full deflection), at

least 3 sieves, collecting vessels and clamping lids must be clamped on

the sieve shaker. If fewer sieves are clamped, it may not be possible to

achieve an amplitude set point of 3 mm.

(Sieves with 200 mm diameter)

Using the device

- 25 -