4.5.1 Adjusting the mains voltage

CAUTION!
Only qualified personnel may change the voltage range on

the device!

CAUTION!
The voltage range may only be adjusted after the mains has

been disconnected.

CAUTION!
If the values 06 or 07 are set at a voltage of 230V~, it will

result in a defect in the sieve shaker during operation.

Failure to observe this will render void the guarantee, and

releases us from liability for any resulting damage to the

device.

1.

Switch main switch to "0".

2.

Switch the device on at the main switch while holding down the

Stop key.
The device is now in "Setup" mode.

3.

The "Sieving Time" display now shows the default setting "04" for

230 V, "06" for 115 V. The value "07" is preset for a mains voltage

of 100 V (in Japan).

4.

These values can be changed with the "Sieving Time" +/- key and

adapted to the mains voltage.

n

230 V - 04

n

115 V - 06

n

100 V - 07

The set value is automatically saved.

5.

After making adjustments, switch on the main switch again.

Installation

- 21 -