2 clamping the torquemaster, Accessories – FRITSCH ANALYSETTE 3 PRO User Manual

Page 36

Advertising
background image

3.

Insert the sieve set and pour in the sieving stock.

4.

Fit the TorqueMaster clamping lid.

5.

Place the clamping unit on the clamping lid. The clamping unit

must be fitted as centrally as possible.

NOTICE!
This symmetry is ensured by correctly adjusting and

securing the toothed belt.

6.

Lightly tighten the toothed belts on both sides and move the

clamping lever upwards, as described in Ä Chapter 6.1.1 ‘Fitting

and clamping the sieves’ on page 23 under point 9.

Make sure that both indicators remain in their central

position.

7.

The clamping system can now be clamped using the cordless

screwdriver.

7.4.2 Clamping the TorqueMaster

Clamping with a cordless screwdriver until it disengages applies a repro-

ducible force of ~ 1 kN to the sieve cover.
Clamping processes during which the friction clutch of the cordless

screwdriver does not disengage (e.g. due to defect, dead battery or

operating error) do not provide the desired clamping force.

The correct clamping force is only achieved if the friction

clutch of the cordless screwdriver triggers during the

clamping process.

The clamping unit must be operated with the defined disengagement

torque of the control element (cordless screwdriver). This disengage-

ment torque must be set by the clamping system manufacturer. Thus

other cordless screwdrivers, which were not set by Fritsch, must not be

used. If cordless screwdrivers with too high a torque are used, it will

cause irreparable damage to the clamping system for which no liability

can be assumed.
The use of cordless screwdrivers that merely have a lower torque will

result in the sieve stack being clamped with an insufficient clamping

force.