7.5.1 Conducting a grinding operation

7.5.1.1

Fitting and clamping the mortar

1.

Move the clamping lever on the belt clamping device downwards

and insert the toothed belt through the holder from the inside to

the outside. (See Ä Chapter 6.1.1 ‘Fitting and clamping the sieves’

on page 23).

2.

Place the mortar and the ball on the vibratory plate. The mortar

must be positioned in the round gap of the vibratory plate without

being wedged.

3.

Unscrew the knurled knob on the toothed belt as far as possible.

4.

Pour in the grinding stock.

5.

Put on the clamping lid.

6.

Attach the knurled knob to the lid

7.

Lightly pull the toothed belt tight and move it upwards until its

teeth interlock with the teeth of the belt clamping device.

8.

Move the clamping lever upwards.

Make sure that the teeth of the toothed belt interlock

with the teeth of the belt clamping device!

9.

By turning the knurled knobs to the right, tighten the mortar

evenly until the toothed belts are tightly clamped!

If the two toothed belts are not evenly clamped, it is

possible that the grinding stock will escape from the

mortar.

Select an amplitude for grinding at which the ball vibrates. The best

grinding results can be achieved at a medium amplitude (1 mm to max. 2

mm), as the impact frequency of the grinding ball is then highest.
Similar to the ANALYSETTE 3 SPARTAN, the amplitude is set manually

with the Plus and Minus key on the control panel and viewed on the lid.

(See Ä Chapter 6.7.1 ‘Displaying the amplitude’ on page 32).

NOTICE!
Do not allow the grinding balls to jump too high. This can

cause damage to the lid!
Start grinding at low a amplitude and increase it slowly - do

not select an amplitude that is too high! (≤ 2 mm)