1 sieving parameters, 2 feed quantity, 3 fitting the micro-precision sieves – FRITSCH ANALYSETTE 3 PRO User Manual

Page 29: Using the device

Advertising
background image

6.4 Micro-precision sieving with the ANALYSETTE 3 PRO

6.4.1 Sieving parameters

Parameters

1 micro sieve

Max. 4 micro sieves

Sieving time

20...30 min

60 min

Amplitude in "MICRO"

operating mode

0.1...0.2 mm

0.5 mm

6.4.2 Feed quantity

Depending on particle size and sieve aperture size, it is possible to sieve

a maximum of 200 mg (up to 1 g for coarse sieves of 50 - 100 µm). Deter-

mine the suitable feed quantity through experiment.

NOTICE!
The device must be in "MICRO" operating mode. The

"MICRO" key must light up, otherwise there is a risk of

destroying the sieves.

In the "MICRO" operating mode the amplitude is limited, so that only

values between 0.1 and 0.5 mm can be selected.

NOTICE!
For sieving with micro-precision sieves, the toothed belts

for dry or wet sieving should be removed from the belt

clamping device. (See Ä Chapter 6.1.1 ‘Fitting and

clamping the sieves’ on page 23).

6.4.3 Fitting the micro-precision sieves

NOTICE!
Clamp a maximum of 4 micro-precision sieves!

Using the device

- 29 -