7.4.4 TorqueMaster malfunctions

For all functions, the two indicators of the clamping unit must be in the

green area. If one or both indicators are in the red area, proceed

according to the following table.

Malfunction

Possible causes

Rectify fault

Both indicators in outer red area

Clamping unit open too much

Open the clamping unit to a lesser

degree, the toothed belt clamp is

possibly too tight and must be loos-

ened by one tooth at the belt

clamping device on both sides.

Clamping unit must be opened far

enough to allow it to be deposited /

clamping unit not attached according

to instructions

Attach according to instructions.

Both indicators in inner red area

respectively

Clamping unit closed too much /

clamping unit not attached according

to instructions

Attach according to instructions.

An indicator in red area on one side

Clamping system attached "very"

asymmetrically

Attach according to instructions.

Make sure that the toothed belt is

fastened symmetrically in the belt

clamping device.

7.4.5 Compatibility

The TorqueMaster must be used with the toothed belt clamping device

(03.3120.00). The design and proper use of this combination will rule out

any damage to the toothed belt.
Generally, the TorqueMaster can also be used with the older belt

clamping device (03.1210.00). However, this combination leads to

increased wear and is therefore not recommended.

7.5 Conversion to vibratory sieve shaker PULVERISETTE 0

The ANALYSETTE 3 can also be used as a vibratory micro

mill PULVERISETTE 0 by using a mortar and the corre‐

sponding grinding ball.

The "vibratory" micro mill PULVERISETTE 0" is used for fine comminution

of dry laboratory sample material or suspensions. For homogenising

emulsions or pastes.