Other settings, Zero point positioning, Bank switching – KEYENCE GV Series User Manual

Page 6

Advertising
background image

6

GV-IM-E

„

1-point area calibration (operation mode: F-2)

1

Press and hold the [SET] button

for at least three seconds with the

workpiece that you want to detect

in place.

2

Release the [SET] button when

"SEt" flashes on the display.

This concludes 1-point area calibration

and the sensor returns to the detection

state.

Configuring the sensitivity setting for reference

surface (DATUM) detection method

„

Reference surface (DATUM) calibration (operation
mode: F-1, F-2)

This method memorizes the state without a workpiece (reference surface) and

use it to perform comparator output when the state differs from the reference

surface (when there is a workpiece).

1

Press the [SET] button once without

a workpiece in place.

2

Press and hold the [SET] button for at

least three seconds without the

workpiece in place.

When performing reference surface calibration, the values are set for slightly above

and slightly below the reference surface. When the detection value falls within this

range, comparator output is turned off. When it falls outside of this range,

comparator output is turned on.

z

Operations during F-1 mode

The setting value can be adjusted by using the [Up] and [Down] arrow buttons,

but the HIGH and LOW setting values cannot be adjusted individually.

z

Operations during F-2 mode

The HIGH and LOW setting values can be adjusted individually by using the [Up]

and [Down] arrow buttons.

• Reference surface calibration cannot be used during the following

states. (Instead, 2-point calibration is performed.)
• The spot reflection indicator (1spot) is flashing on the sensor head or

sensor amplifier.

• The setting is control output 2.
• A mode other than F-1 mode or F-2 mode is being used.

• Reference surface calibration can only be set when the operation mode

is F-1 or F-2.

• When performing reference surface calibration, the current value is set

to 0.

„

Clearing reference surface (DATUM) detection
method

When the reference surface (DATUM) detection indicator is lit, press and hold

the [Up] and [Down] arrow buttons for at least three seconds. The reference

surface (DATUM) detection indicator turns off and the sensor returns to

distance detection method.

The value is canceled if a form of calibration other than reference surface

calibration.

Zero point positioning

Sets the current value to zero (shift target value).

1

Press the [Up] and [Down] arrow buttons simultaneously without a

workpiece in place.

The current value becomes "0" and zero point positioning is complete.

When the shift target value is set, performing zero point positioning does not
make the current value "0". Instead, it becomes the set value for the shift
target value.
For more information about setting the shift target value, see page 9 of this
Instruction Manual.

Pressing the [Up] and [Down] arrow buttons simultaneously for at least

three seconds cancels the zero point (shift target value).

Initial reset (initialization) and custom save function

„

Initial reset (initialization)

Returns all of the settings to the factory defaults.

1

While pressing the [MODE] button, press the [SET] button five times.

2

Press the [Down] arrow button to display "rSt no".

When "rSt no" is displayed, pressing the [MODE] button returns to the

main screen without initializing.

3

Press the [Up] or [Down] arrow button to select "rSt ini", and then

press the [MODE] button to initialize.

„

Custom save (Saving the settings)

Saves all of the settings. The saved settings can be loaded at a later time.

1

While pressing the [MODE] button, press the [SET] button five times.

2

Press the [Left] or [Right] arrow button to display "SAv".

3

Press the [Down] arrow button to display "SAv no".

When "SAv no" is displayed, pressing the [MODE] button returns to the

main screen without saving the settings.

4

Press the [Up] or [Down] arrow button to select "Sav YES", and then

press the [MODE] button to save the settings.

Performing custom save overwrites the settings from the previous custom
save.

„

Custom reset (Loading the settings)

Loads the settings saved with custom save.

1

While pressing the [MODE] button, press the [SET] button five times.

2

Press the [Down] arrow button to display "rSt no".

When "rSt no" is displayed, pressing the [MODE] button returns to the

main screen without performing custom reset.

3

Press the [Up] or [Down] arrow button to select "rSt CSt", and then

press the [MODE] button to load the settings.

Performing custom reset erases the current settings.

Bank switching

The bank function allows two patterns of sensitivity settings to be saved in channels

1 and 2. (This is useful during operations such as switching the setup.)

The bank switching function can only be used when the external input is set to "bnk" in

the OPTIONAL settings.

Hold down the [MODE] button during main screen and press the [Up] arrow button to

switch banks, which allows the settings to be configured for each bank.

During the keylock state, the banks are switched with external input.

When the external input is off, bank A is used. When the external input is on,
bank B is used.

Other Settings

HIGH setting value

LOW setting value

Set the values for the

upper and lower

distance.

Note

Note

Reference

Reference

Reference

Note

Reference

Note

Note

DATUM

DATUM

BANK

BANK

SET

MODE

1
2

SET

MODE

1
2

MODE

Bank A is in use : Bank indicator is off

Bank B is in use : Bank indicator is on