Installation environment

To use this product properly and safely, avoid installing it in the following locations. Doing so

may lead to product breakdown.

• Location that is humid, dusty or poorly ventilated

• Location where the temperature becomes high, such as a place exposed to direct

sunlight

• Location where there are flammable or corrosive gases

• Location where the product may be directly subjected to vibration or impact

• Location where water, oil or chemicals may splash onto the product

Considerations for improving noise immunity

• Do not install in enclosures where high voltage devices are installed.

• Locate as far away as possible from power lines.

• Locate as far away as possible from devices that generate intense electric or magnetic

fields such as solenoids and chopper circuits.

• Use separate conduit for I/O lines, power lines, and high voltage lines to prevent noise-

induced erroneous operation.

• Ground protective earth terminals and shield wires to a ground resistance of 100 ohms

or less.

Precautions on CE Marking

This device conforms to CE marking directives when the following conditions are met:

Make sure these conditions are met if using this device within the EU nations.

The applicable standards (EMC Directive) are listed below:

EMI:EN61326-1, Class A

EMS:EN61326-1

Limit the length of the power supply cable and all input/output cables that are connected to

the terminal panel of the controller to 30 m or less.