Shift adjustment function

Adjusts the analog value zero point using the UP/DOWN keys for shift
adjustment.

Holding down the UP/DOWN keys to change the shift value faster.
Pressing the ZERO RESET key for 2 seconds cancels the shift adjust-
ment value, and 0 V (12 mA) is output when the target is at the reference
position.

Shift adjustment range
LK-031/036/081/086

LK-501/503

LK-501/503 (Long-range mode)

(High-precision mode)

Data in ( )* applies to the LK-081.
Data in ( )** applies to the LK-501/503 in long-range mode.

FUNCTIONS SHIFT/SPAN

Span adjustment function

Adjusts the analog value inclination using the UP/DOWN keys for span
adjustment. Use the span adjustment when the sensor head is tilted or
the target surface condition affects the analog output characteristics.

Holding down the UP/DOWN keys to change the span value faster.
Pressing the RESET key for 2 seconds cancels the span adjustment
value, and the output characteristics are reset to the factory-set values.

Span adjustment range
LK-031/036/081/086

LK-501/503

LK-501/503 (Long-range mode)

(High-precision mode)

Data in ( )* applies to the LK-081.
Data in ( )** applies to the LK-501/503 in long-range mode.

ZERO

RESET

LK-031

30.015 mm

30.015 mm

LK-031

V

V

10

5

0

-5

-5

(-15)*

(-250)**

5

(15)*

(250)**

-10

12

Distance (mm)

Output voltage

(V)

RESET

LK-031

V

LK-031

V

2.000 mm

V

10

5

0.5

0

-5

-0.5

-5

(-15)*

(-250)**

5

(15)*

(250)**

-10

12

Distance (mm)

Output voltage

(V)

Value is inaccurate due to reflective property of the target surface.

10

5

10

-100

100

-10

12

Distance (mm)

Output voltage

(V)

10

1

0

-1

50

-50

100

-100

-10

12

Distance (mm)

Output voltage

(V)