Timing diagram

• Laser emission only starts approximately 2 seconds after the laser

emission stop input terminal is connected.

• While the timing input is connected, laser emission stops. The analog

output will hold the last value before the timing input signal turns on.

INPUT CHARACTERISTICS

Minimum input time

Data in ( )* applies to the LK-081.
Data in ( )** applies to the LK-501/503.

Synchronous (timing) input

Note: The analog output is retained for the following amount of time after
disconnection of the synchronous input terminal.

LK-031/036: Approx. 10 ms
LK-081/086: Approx. 20 ms
LK-501/503 (Sensitivity: AUTO): Approx. 100 ms
LK-501/503 (Sensitivity: HI or LOW): Approx. 16 ms

AUTO ZERO input

Note: The analog output is retained approximately 10 ms after the
synchronous input terminal is disconnected.

12 V

0 V

Synchronous

(timing) input

Analog output

Laser emission

stop input

ON

OFF

ON

OFF

ON

OFF

AUTO ZERO

input

2ms

(2.5 ms)*

(4.5 ms)**

10ms

(20 ms)*

OFF

ON

Sensitivity: AUTO: 100 ms

Sensitivity: HI or LO: 16 ms

2ms

(2.5 ms)*

(4.5 ms)**

2ms

(2.5 ms)*

(4.5 ms)**

OFF

ON