Chapter 4. L-Cell Installation on Horizontal Beams

4-3

L-Cell Mounting Methods

The sensors are mounted at the neutral axis on the web of the beam, at a
45-degree angle, halfway between the loading point and the support point.
The following methods provide varied degrees of accuracy.

I. Same Side: One L-Cell to the left and one L-Cell to the right of the load
point equidistant between the load and the support. The L-Cells are
mounted on the same side of the web. The excitation of the L-Cell to the
right of the load point must be reverse wired (See page 4-8). This arrange-
ment provides average accuracy performance.

II. Opposite Side: One L-Cell to the left and one L-Cell to the right of the
load point equidistant between the load and the support. The L-Cells are
mounted on opposite sides of the web. This arrangement provides average
accuracy performance.

III. Back-to Back: Two L-Cells mounted back-to-back on the web. They can
be mounted beck-to-back on either side of the load point, equidistant
between the load and the support. This arrangement provides average
accuracy performance.

Figure 4-3. Placement of Two L-Cells on same side of web

Load Point

Top of L-Cell Point
Toward LoadPoint

A

B

Reverse
wired

Figure 4-4. Placement of Two L-Cells on opposites of web

Load Point

Top of L-Cell Point
Toward LoadPoint

A

B

Opposite
side of
beam

Column

(or beam)

Supporting

Instrumented

Beam

Load Point from

Vessel or End of

Diagonal Beam

Two L-Cells

back-to-back

Horizontal or

Diagonal Beam

1

/

2

L

L = Distance from Load Point to Column or Beam Support

Minimum of 18” (457mm)

or beam height

Figure 4-5. Placement of Two L-Cells back-to-back