Obstructions, Aluminum) – Kistler-Morse KM L-Cell Bolt-On User Manual

Page 47

Advertising
background image

Chapter 5. L-Cell Installation on Skirt-Supported Vessels

5-10

See Figure 5-14. Minor obstructions may occur near the top or bottom of
a tier. The effective tier height is the distance between the edge of the tier
and the minor obstruction. If the effective tier height is 2.6’ (0.8m) or
greater, modify the vertical location of the L-Cell set so the set is at the
center of the effective tier height. If the effective tier height is less than
2.6’ (0.8m), relocate the L-Cell set as shown in Figure 5-15.

Figure 5-14. Modifying L-Cell Vertical

Locations because of Minor Obstructions

Near Tier Edge

Note: Minimum effective tier height is 2.6’ (0.8m).
If effective height is less than 2.6’ (0.8m), relocate
L-Cell set to another location.

Modified Location to
Center L-Cells within
Effective Tier Height
(at this point only)

3.5’

(1.1m)

12

12

12

12

12

12

12

12

Effective

Tier

Height

7.0’ (2.1m)

2” to 4”
(51mm to 102mm) Diameter
Pneumatic Tubes through Tier

4.0’
(1.2m)

Tier

Height

8.0’

(2.4m)

‘Ideal’ Locations

Mount L-Cells
1.3’ (0.4m) from
Major Obstruction

Figure 5-12. Mounting L-Cells in Alternate
Locations because of Major Obstructions

14’-9” (4.5m)

Diameter

Typical Layout

14’-9” (4.5m) Diameter

Modified Layout because

of Major Obstruction

Do not install L-Cells within 1.3’ (0.4m) of major
obstruction. Install as shown on Modified layout below.

Major Obstruction

at Ideal Location
for an L-Cell Set

Mount L-Cells
1.3’ (0.4m)
from Major
Obstruction

Figure 5-15. Mounting L-Cells on Second Tier

First

Tier

Second

Tier

Must have
enough room to
work inside

2.6’ (0.8m) Minimum
Height between
Seam Welds or
Bolt Lines

Center L-Cells
Vertically between
Seam Welds or
Bolt Lines

Many Obstructions

in First Tier

Cone

Obstructions can be classified as major or minor:

Figure 5-12 shows an example of a major obstruction. Do not place an
L-Cell set within 1.3’ (0.4m) of a major obstruction.

Figure 5-13 shows an example of a minor obstruction. Do not place an
L-Cell set within 0.7’ (0.2m) of a minor obstruction — less than 4”
(102mm) in size.

Figure 5-13. Mounting L-Cells in Alternate
Locations because of Minor Obstructions

Maintain 0.7’ (0.2m) Minimum
Distance between L-Cells and
Minor Obstruction (seam weld)

Modified Location to

Avoid Placement

of L-Cells near

Minor Obstruction

‘Ideal’ Locations, based
on Symmetrical Spacing

Seam
Weld

r

u

c

t

Obstructions

(Aluminum)