Kistler-Morse KM L-Cell Bolt-On User Manual
Page 29
Chapter 4. L-Cell Installation on Horizontal Beams
4-4
Figure 4-7. Placement of Four L-Cells on a Beam
Column
(or beam)
Supporting
Instrumented
Beam
(typical)
Load Point
from Vessel
L
Minimum of 18” (457mm)
or beam height
1
/
2
L
1
/
2
L
L
Minimum of 18” (457mm)
or beam height
Figure 4-6. Placement of single L-Cell Set to Left of Load Point
Column
(or beam)
Supporting
Instrumented
Beam
Load Point from
Vessel or End of
Diagonal Beam
L = Distance from Load Point to Column or Beam Support
Minimum of 18” (457mm) or
beam height
1
/
2
L
Top of L-Cell
Points Toward
Load Point
Horizontal or
Diagonal Beam
A
B
IV. Single L-Cell: One L-Cell mounted on the web. It can be located on either side of
the web and either side of the load point. It should be mounted equidistant between the
load and the support. This arrangement yields the lowest accuracy performance.
V. Four L-Cells: Four L-Cells mounted back-to-back on the web. One pair mounted
back-to-back to the right of the load point and one pair mounted back-to-back to the
left of the load point. The pairs are mounted equidistant between the load and the
support. This arrangement provides a degree of accuracy improvement over a two
L-Cell arrangement.