Manual calibration – Kistler-Morse KM L-Cell Bolt-On User Manual

Page 77

Advertising
background image

Appendix D. System Calibration

D-3

Figure D-2. Manual Calibration Line

Weight Signal in Counts

Indicated
Weight

live load =

Zero_Cal

Scale Factor
Weight

Scale Factor
Counts

Note: Scale Factor Counts and Scale Factor Weight
establish the slope of the Manual calibration line.
Zero_Cal establishes the location of the line.

Slope =

Scale Factor Weight

Scale Factor Counts

See Figure D-2:

1.

Refer to the signal processor manual to determine how to obtain the
A/D converter sensitivity, expressed in Counts/mV. Record this value.

2.

Record the L-Cell sensitivity (S):

Carbon steel vessel
35 mV/1000 psi (35 mV/0.7kg/mm

2

)

Aluminum vessel
80 mV/1000 psi (80 mV/0.7kg/mm

2

)

3.

Refer to the Application Data Form for the vessel (if you cannot locate
the completed form, contact K-M for a copy). Record the maximum live
load and the stress.

4.

Record the current live load in the vessel.

5.

Calculate the Manual calibration values:
Scale Factor Weight = Maximum live load
Scale Factor Counts = S x Counts/mV x Stress
Zero_Cal = Current live load

6.

Refer to the signal processor manual to input the calibration values.

Example: You are using L-Cells to monitor a 15’ diameter carbon steel, skirt-
supported vessel with 0.187” wall thickness. The vessel currently contains
50,000 lbs of material and can hold a maximum of 200,000 lbs. Following the
procedure:
1.

Counts/mV = 699.05 (from signal processor)

2.

S = 35 mV/1000 psi (from L-Cell manual, for carbon steel vessel) = 0.035
mV/psi

3.

From the Application Data Form, the maximum live load is 200,000 lbs.
The stress is:

200,000 lbs

3.14 x (15’ x 12”/ft) x 0.187”

=1892 psi

4.

Current live load = 50,000 lbs

5.

Calculate the values for the calibration:
Scale Factor Weight = Maximum live load = 200,000 lbs
Scale Factor Counts = S x Counts/mV x Stress

= .035 mV/psi x 699.05 Cnts/mV x 1892 psi
= 46,291 Counts

Zero_Cal = Current live load = 50,000 lbs

Manual Calibration