3 troubleshooting – Intek RheoVac 950A User Manual

Page 22

Advertising
background image

- 20 -

4.3 TROUBLESHOOTING

4.3.1 Diagnostic

Messages

Table 6 provides a guide for plant personnel to identify causes of problems and determine
appropriate actions to resolve problems observed. If problems are encountered and factory
assistance is desired, please contact the factory.

Table 6 Troubleshooting Guide (Diagnostic Messages)

MESSAGE CODE

Description/Symptom

PROBABLE CAUSE

ACTION

MSG-0

Communications not

being received from

probe(s)

1.

Improper cable hookup

2.

Blown main fuse

3. Failed

RS-485

circuit

4.

Damaged flow sensor

1.

Verify plug-in connector is properly
mated

2.

Check F1 fuse on probe board

3.

Check all cable connections

4. Contact

factory

MSG-1

Invalid data received

from probe(s)

1. Failed

RS-485

communications

component

1. Check wiring
2. Contact

factory

MSG-2

Flow sensor heater

“OFF”

1. Blown heater fuse
2.

Failed electronic component

1. Contact

factory

MSG-3

RS sensor power

“OFF”

1. Probe temperature too high
2.

Liquid water on probe tips

1. Check that probe temperature is

<160ºF

2. Contact

factory

MSG-4

RS heater “OFF”

1. Component

failure

1. Contact

factory

MSG-5

Circuit issue

1.

Problem with circuitry

1.

Contact factory

MSG-6

Temperature alarm

(above 210ºF/99ºC)

1.

Steam in exhaust pipe

1. Remove probe or cool line ASAP!
2.

Once line has cooled down and
probe is reinstalled, check unit for
proper function

MSG-7

Wet probe

1.

Liquid water on probe tip

1. Remove

probe

ASAP!

2. Contact

factory

MSG-8

RS sensor problem

1. RS

sensor

problem

1.

Remove from line, allow 24 hrs with
power on to dry out RS sensor,
reinsert probe

2. Contact

factory

Mass flow output

saturates high, will not

respond to flow changes

1. Flow rate is not within range of

calibration

2. Blown heater fuse
3.

Failed electronic component

1. Contact factory about re-ranging

instrument

2. Contact

factory

Mass flow output

saturates low, will not

respond to flow changes

1. Flow rate is not within range of

calibration

2.

Failed electronic component

1. Contact factory about re-ranging

instrument

2. Contact

factory