Intek RheoVac 950A User Manual

Page 43

Advertising
background image

-C6-

I:\OFFICE\WPMANUAL\RV950A Appendix C revB.doc

Feb. 4, 2010

Table C2.1: Modbus Server Global Settings

Address Description

Allowed

Values Default

Format

Read

Only?

0001

Restart Modbus Server Write ‘1’ to restart 0

16-bit

unsigned

No

0002

RheoVac Serial Port

Set to ‘0’ internally by
RheoVac System

0

16-bit unsigned

Yes

0003 Baud

Rate

2400,4800,9600,14400,
19200 bps

9600 16-bit

unsigned

No

0004

Data Bits

7 or 8

8

16-bit unsigned

No

0005

Stop Bits

0 = 1 stop bit
1 = 1.5 stop bits
2 – 2 stop bits

2

16-bit unsigned

No

0006

Parity

0 = No Parity
1 = Odd Parity
2 = Even Parity
3 = Mark Parity
4 = Space Parity

0

16-bit unsigned

No

0007

Modbus Address

1 – 255

1

16-bit unsigned

No

0008

Communication Mode

0 = ASCII
1 = RTU

0

16-bit unsigned

No

0009 Conversion

Mode

0 = Multiplied Integer
1 = Single Float

0

16-bit unsigned

No

0010

Number of Probes

1 – 3

1

16-bit unsigned

Yes

C.2.3

MODBUS DATA SETTINGS (Multiplied Integer Mode)

The values listed in Table C2.2 are used only when the Modbus server is in “Multiplied Integer” mode
(“0” written to register 09). These register values determine the power of 10 that is used to multiply each
parameter (effectively shifting the decimal point) before it is transmitted as a 16-bit unsigned integer.

Table C2.2: Modbus Data Multiplier Registers

Address Description Allowed

Values

Default Read

Only?

0011 Multiplier

power for Actual Volume Flow

0 – 3

16-bit unsigned

No

0012

Multiplier power for Total Mass Flow

0 – 3

16-bit unsigned

No

0013

Multiplier power for Water Vapor Flow

0 – 3

16-bit unsigned

No

0014

Multiplier power for RheoVac Pressure

0 – 3

16-bit unsigned

No

0015

Multiplier power for Water Vapor Specific Volume

0 – 3

16-bit unsigned

No

0016

Multiplier power for Water Vapor to Air Mass Ratio

0 – 3

16-bit unsigned

No

0017

Multiplier power for Relative Saturation

0 – 3 16-bit

unsigned No

0018

Multiplied power for Water Vapor Partial Pressure

0 – 3

16-bit unsigned

No

0019

Multiplier power for Air Flow In-Leak

0 – 3

16-bit unsigned

No

0020

Multiplier power for Probe Temperature

0 – 3 16-bit

unsigned No

C.2.4

MODBUS DATA REGISTERS (Multiplied Integer Mode)

When in Multiplied Integer mode, the RheoVac 950 data values are multiplied by a power of ten specified
in the registers listed in Table C2.2. The resulting values are then transmitted as 16-bit unsigned
integers. Each value therefore needs only 1 register (2 bytes) of allocated space. In this mode, the data
registers are allocated in pairs, with each data value followed by the multiplication power corresponding
to it. For example, assuming all multiplier registers are set to 3, the RheoVac Pressure is transmitted as
follows:

RheoVac Pressure (probe reading) = 1.257
Register 117 (RheoVac Pressure) = 1257
Register 118 (Multiplier Value) = 3