Table C2.4: Modbus Data Registers (Single Float Mode)

Address Description

Format

Read Only?

101 – 105

Probe Serial Number

10 ASCII characters

Yes

106

Error Code: see Section C.2.6

16-bit unsigned

Yes

111 – 112

Actual Volume Flow Single-Precision

Float

Yes

113 – 114

Total Mass Flow

Single-Precision Float

Yes

115 – 116

Water Vapor Flow

Single-Precision Float

Yes

117 – 118

RheoVac Pressure

Single-Precision Float

Yes

119 – 120

Water Vapor Specific Volume

Single-Precision Float

Yes

121 – 122

Water Vapor to Air Mass Ratio

Single-Precision Float

Yes

123 – 124

Relative Saturation

Single-Precision Float

Yes

125 – 126

Water Vapor Partial Pressure Single-Precision

Float

Yes

127 – 128

Air Flow In-Leak

Single-Precision Float

Yes

129 – 130

Probe Temperature

Single-Precision Float

Yes

C.2.6

RHEOVAC INSTRUMENT ERROR CODE

The error code address will return 4 ASCII characters (ASCII Mode) or 2 Hex bytes (RTU Mode) that
represent a binary number. A value of 1 in the binary representation will indicate a specific error
depending on the location of the 1 in the binary number as described in Table C.2.5. Examples are
shown in Table C.2.6.

Table C2.5: Modbus Error Code Decryption Table

Error

Indication CalX MSG8 MSG7 MSG6 MSG5 MSG4 MSG3 MSG2 MSG1 MSG0

Binary Value

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

0 or 1

Table C2.6: Modbus Error Examples

ASCII Mode Output

(data portion only)

RTU Mode Output

(data portion only)

Binary Representation

Indicated Error Messages

30 30 30 30

00 00

0000 0000 0000 0000

No errors

30 30 30 31

00 01

0000 0000 0000 0001

MSG0

30 30 34 43

00 4C

0000 0000 0100 1100

MSG6, MSG3, MSG2

30 31 43 41

01 CA

0000 0001 1100 1010

MSG8, MSG7, MSG6, MSG3, MSG1

30 33 30 34

03 04

0000 0011 0000 0100

CalX, MSG8, MSG2

C.2.7

MODIFYING THE MODBUS CONFIGURATION FILE

To modify the configuration file, the RheoVac Model 950 must be attached to a local network. Once
connected, Windows Explorer can be used to browse into the RheoVac Model 950 similar to a PC. Using
the name of the RheoVac Model 950 as the Network ID, the Modbus settings file can be found at the
following location:

\\<NETWORK_ID>\RheoVac\init\serial.ini

Figure C2.1 shows a sample serial.ini file. The Modbus settings are listed under the section titled
“[Modbus]”. In addition, when the Modbus “Mode” setting is 0 (Multiplied Integer), the settings listed
under the “[DecLoc]” section define the decimal location for each of the 10 process variables. These
values can be changed to fit the needs of the existing serial Modbus network. Once the changes have
been made and the file has been saved in its current location, the RheoVac Model 950 must be rebooted