To allow for future expansion and up to three different probes, each probe is allocated a space of 100
registers, with the registers for the first probe beginning at address 0101. Registers for the second probe
begin at address 0201, and the third probe begins at register address 0301. Table C2.3 details the
RheoVac Model 950 data register setup when in Multiplier Integer mode.

Table C2.3: Modbus Data Registers (Multiplied Integer Mode)

Address Description

Format

Read Only?

0101 – 0105 Probe Serial Number

10 ASCII characters

Yes

0106

Error Code: see Section C.2.6

16-bit unsigned

Yes

0111 Actual

Volume

Flow 16-bit

unsigned

Yes

0112 Multiplier

Value

16-bit unsigned

Yes

0113

Total Mass Flow

16-bit unsigned

Yes

0114 Multiplier

Value

16-bit unsigned

Yes

0115

Water Vapor Flow

16-bit unsigned

Yes

0116 Multiplier

Value

16-bit unsigned

Yes

0117

RheoVac Pressure

16-bit unsigned

Yes

0118 Multiplier

Value

16-bit unsigned

Yes

0119

Water Vapor Specific Volume

16-bit unsigned

Yes

0120 Multiplier

Value

16-bit unsigned

Yes

0121

Water Vapor to Air Mass Ratio

16-bit unsigned

Yes

0122 Multiplier

Value

16-bit unsigned

Yes

0123 Relative

Saturation 16-bit

unsigned

Yes

0124 Multiplier

Value

16-bit unsigned

Yes

0125

Water Vapor Partial Pressure 16-bit

unsigned Yes

0126 Multiplier

Value

16-bit unsigned

Yes

0127

Air Flow In-Leak

16-bit unsigned

Yes

0128 Multiplier

Value

16-bit unsigned

Yes

0129 Probe

Temperature 16-bit

unsigned

Yes

0130 Multiplier

Value

16-bit unsigned

Yes

C.2.5

MODBUS DATA REGISTERS (Single Float Mode)

In Single Float mode, all of the data values are converted into the 4-byte, big-endian hexadecimal string
corresponding to their single-precision float representation. The Modbus client program must be capable
of converting these 4-byte values back into single-precision float values. If the client software is unable
to make this conversion, the Multiplied Integer mode must be used. Data in the Single Float mode is
transmitted as follows:

RheoVac Pressure (probe reading) = 1.257
Single-Float Hex Representation = 0x3FA0E560
Register 117 (high word) = 0x3FA0
Register 118 (low word) = 0xE560

As in the Multiplied Integer mode, each probe is allocated a space of 100 registers, with the registers for
the first probe beginning at address 0101. Registers for the second probe begin at address 0201, and the
third probe begins at register address 0301. Table C2.4 details the RheoVac Model 950 data register
setup when in Single Float mode.