Technical data – VACUUBRAND MZ 2C EX +IK+EK User Manual

Page 16

Advertising
background image

page 16 of 75

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

Technical data

Type

MZ 2C EX

MZ 2C EX

+ AK + EK

MZ 2C EX

+ IK + EK

MD 4C EX

MD 4C EX

+ AK + EK

MV 10C EX

MV 10C EX

+ AK + EK

ATEX approval

Inner part (pumped gases)

Outer part

Surrounding of the pump with inert gas purge

Surrounding of the pump without inert gas purge

Motor

II 2G IIC T3 X

II 2G IIB T4 X

II 3G IIB T4 X

see motor rating plate

Max. pumping speed* according to

ISO 21360

m

3

/h

1.9

3.7

8.1

Ultimate vacuum* (absolute) without

gas ballast

mbar

12

3

2

Ultimate vacuum* (absolute) with

gas ballast

mbar

18

10

10

Max. permitted pressure at the inlet

and the outlet (absolute)

bar

1.1

Max. permitted pressure difference

between inlet and outlet

bar

1.1

Max. permitted pressure at inert gas

connection (absolute)

bar

1.1

Max. permitted ambient temperature

during operation

°C

+10 to +40

Max. permitted ambient temperature

during storage

°C

-10 to +60

Permitted gas inlet temperature

°C

+10 to +40

Max. surface temperature of the

pump (outer part)

°C

110

Max. permitted atmospheric moisture

during operation

(no condensation)

%

30 to 85

Degree of protection of the pump IEC 529

Pump

Pump + AK + EK / Pump + IK + EK

IP 54**

IP 52**

IP 54**

IP 54**

IP 54**

IP 54**

Nominal current draw

A

1.2

2

2 x 2

Maximum start-up current /

Start-up duration

5.5 A /

125 ms

10.2 A /

125 ms

2 x 10.2 A /

125 ms

Max. permitted range of voltage sup-

ply / frequency

230V ± 10% / 50 Hz

Integral leak rate

mbar*l/s

0.1

We reserve the right for technical modifications without prior notice!

*

Technical data according to EN 61010-1 and EN 1012-2 recommendation. The pump achieves its ultimate pumping speed and ultimate vacuum only at operating

temperature (after approx. 15 min.).

** Only when positioning the pump horizontal.