• All connection lines at the pump (e. g. inlet, outlet, gas ballast connection, inert

gas connection) have to be performed conductive grounded.

• Connect pump and all connected components to ground.

• Avoid contact of the pump with rusty metal parts, e. g. by positioning the pump

with sufficient distance to surrounding parts.

• The pump is designated for a low degree of mechanical stress. Protect the pump

from damage and beats. In case of damage switch off the pump immediately,

check leak rate if necessary.

• Connection lines at the pump inlet have to be gas tight. Particles and dust must

not be aspirated, the user has to provide appropriate filters if necessary. The user

must ensure their suitability concerning gas flow, chemical resistance and safe-

ness against clogging prior to use.

• Connect an exhaust line gas tight at the pump outlet if necessary. Always dispose

of exhaust gases appropriately (e.g. into a fume hood). If there is risk of release of

dangerous or polluting fluids, install an appropriate system to catch and dispose

of those fluids.

• Reduce the transmission of vibration and prevent mechanical load due to rig-

id pipelines. Insert elastic hoses or flexible elements as couplings between the

pump and rigid pipes.

Attention: Flexible elements tend to shrink when evacu-

ated.

• Use of a suitable valve to isolate the pump from the vacuum system is recom-

mended to allow the pump to warm up before pumping condensable vapours or

to clean the pump before it is switched off.

• The gas outlet must never be blocked. The exhaust line has always to be free

(pressureless) to ensure an unimpeded discharge of gas.

• Especially if the gas ballast valve is open, a power failure may cause accidental

ventilation of the pump. In case this constitutes a potential source of danger, take

appropriate safety measures.

• Make sure ventilation is adequate especially if the pump is installed in a housing

or if the ambient temperature is elevated. Provide external ventilation if neces-

sary. Keep a distance of minimum 20 cm between fan and ambient parts.

Avoid throttling losses by using connecting pipes with large diameter and by keeping

them as short as possible.

In case of perturbing exhaust noise connect an exhaust hose. Install outlet pipelines

always falling to avoid backflow of condensate towards the pump.

When assembling, ensure

vacuum-tightness. After assembly, check the whole sys-

tem for leaks. Secure hose connections at the pump appropriately against accidental

detaching.

Attach the pipelines of the coolant circuit to the respective hose nozzles (hose noz-

zles 6-8 mm, see image) at the vapour condenser. Check hose connections prior to

starting operation of the cooling system.

Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their ac-

cidental slipping.

Use and operation

Installing in a vacuum system

NOTICE

NOTICE