Notice – VACUUBRAND MZ 2C EX +IK+EK User Manual

Page 7

Advertising
background image

page 7 of 75

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

The inner (pumped media) and outer part of the pump have to be evaluated

separately because of their different conformity (see also section ”Technical

data” and ”Notes regarding devices with ATEX conformity”).

The pumps are not suitable for pumping dust and have no approval for operation

below ground.

The pumps are not suitable for pumping self inflammable substances, for substanc-

es which are inflammable without air and explosive substances.

Electrical and vacuum components might require a separate approval.

Include the components in the recurrent tests of tightness for pressure vessels, in

the case of hits and beats an intermediate leak test is required.

Due to the high compression ratio of the pumps, pressure at the outlet port might

be generated being higher than the max. permitted pressure compatible with the

mechanical stability of the system.

Always provide a free and pressureless exhaust pipeline.

Do not permit any

uncontrolled pressurizing (e.g. make sure that the exhaust

pipeline cannot become blocked). If there is an exhaust isolation valve, make sure

that you cannot operate the equipment with the valve closed.

Risk of bursting!

• Secure coolant hoses at the hose nozzles (e.g. with hose clip) to prevent their

accidental slipping.

• Comply with

maximum permissible pressures at inlet and outlet and pressure

differences, see section ”Technical data”. Do not operate the pump with overpres-

sure at the inlet.

Attention: Flexible elements tend to shrink when evacuated.

Provide a firm level platform for the equipment and check that the system to be

evacuated is mechanically stable and that all fittings are secure. Ensure a stable po-

sition of the pump without any mechanical contact except of the pump feet. Comply

with all applicable safety regulations.

Pumping units (pumps with vapour condenser) have to be installed protected

against hits from outside and protected against splinters (against implosion).

Adopt suitable measures prior to the first use.

Obey all relevant requirements (regulations and guidelines) and adopt suitable safe-

ty measures.

Pay attention to the max. permitted ambient temperature and make sure

ventilation is adequate especially if pump is installed in a housing or if

ambient temperature is elevated, install an external automatic ventilation

system if necessary.

If pumping hot process gases make sure that the maximum permitted gas inlet

temperature is not exceeded.

If the equipment is brought from cold environment into a room for operation,

allow the equipment to warm up (pay attention to water condensation on cold

surfaces).


The work clothing of the user must not lead to electrostatic discharge which may

ignite the explosive atmosphere.

NOTICE

Advertising
This manual is related to the following products: