The pump achieves its pumping speed, ultimate total vacuum and vapour pumping

rate only at operating temperature (after approx. 15 minutes).

Prevent internal condensation, transfer of liquids or dust. The diaphragm and

valves will be damaged, if liquids are pumped in significant amounts.

Let the pump run with gas ballast to reduce condensation of pumped substances

(water vapour, solvents, ....) in the pump.

terminal cover

(at the pump motor)

switch for resetting

the motor protection

motor capacitor

The motor is protected by a thermal cutout with manual reset combined with an

overcurrent protection.

Attention: Only manual reset is possible. Determine cause of switching off and

eliminate. Allow sufficient cooling of the pump prior to switching on again (approx.

5 minutes).

Reset of the pump must performed only by a suitably trained technician!

Separate the pump from mains and wait two minutes before opening the terminal

until hot parts have discharged.

Attention: Open the terminal only outside of explosive atmospheres and only by

suitably trained personnel.

Unscrew the terminal cover from the pump motor.

Push the switch inside the terminal.

Screw the terminal cover correctly, screw tightening torque: 23 Nm.

Attention:

Absolutely check the motor capacitor for leaks and capacity if the motor pro-

tection has been activated.

Check the complete pump for leaks in the gas line (integral leak rate < 0.1 mbar

l / s)!