To control the faultless function of the diaphragm it could be recommendable to

install a gas specific detector at the outlet of the inert gas purge. A signal from

the detector indicates a diaphragm crack. Switch off the pump immediately and

check diaphragms.

If the surrounding of the pump is not an explosive atmosphere it is possible to

leave out the inert gas purge.

Connection of the inert gas gas ballast

Connect the inert gas connections at the gas ballast and the crankcase not serial

but parallel.

Use only inert gas at the gas ballast connection. Therefore replace the blind cap

at the gas ballast connection by the hose nozzle enclosed.

Gas ballast connection lines (inlet and outlet) have to be performed conductive

grounded and gas tight (designation of the pump connections see figures).

Provide an external pressure limitation to a pressure of 1.1 bar absolute if inert

gas is connected to the pump or at an air admittance valve. The inert gas has to

be dry and pure. Provide an appropriate control system to supervise the inert gas

flow which switches off the pump in case of failure (e. g. absence of inert gas) if

necessary.

Attention: Important notes regarding the use of gas ballast

When using air rather than inert gas, risk of significant damage to equipment and/

or facilities, risk of personal injury or even loss of life exists due to the formation

of hazardous and/or explosive mixtures if air and pumped media react inside or

at the outlet of the pump.

Outlet of the pump

Always connect an exhaust lines at the pump outlet (small flange or hose nozzle

10 mm). Perform the outlet line conductive grounded and gas tight (designation

of the pump connections see figures). Always connect the exhaust to a suitable

treatment plant (e. g. hood).

Potential explosive mixtures at the outlet of the pump have to be drained appro-

priately, sucked off or diluted with inert gas not explosive mixtures.

If there is a danger of deposits in the pump chamber (check inlet and outlet of the

pump) control the pump chamber regularly and clean if necessary.

If there is a risk of ignition of gases or gas mixtures before or behind the pump the

user has to provide suitable safety devices against incoming of flames according

to EN 12874. The user must ensure the suitability concerning gas flow, chemical

resistance and safeness against clogging prior to use.

Overpressure safety valve at the outlet of the pump

The overpressure safety valve at the outlet prevents reliable high pressures in the

system. Check overpressure safety valve regularly and replace if necessary.

Attention: If the exhaust pipeline is blocked the pumped gases may escape into

the environment through the overpressure safety valve. If necessary take appro-

priate measures.

If pumping dangerous gases install a second completely independent exhaust

pipeline instead of the overpressure safety valve. If necessary assemble a hose

nozzle (thread G 1/4) instead of the overpressure safety valve to lead away the

gases or condensate through a second exhaust pipeline. Attention: Never com-

bine the exhaust pipelines, i. e. provide a separate line for each exhaust.

Attention:

Deposits and condensate may lead to increased temperature and

to excess of the max. permitted temperatures! Increased temperatures possibly

may lead to ignition of flammable mixtures inside the pump.

Use inert gas gas ballast if necessary and check pump regularly for deposits.

Install a gas washing bottle, condenser, filter, separator, etc. at the inlet of the

pump if necessary.