VACUUBRAND MZ 2C EX +IK+EK User Manual

Page 44

Advertising
background image

page 44 of 75

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

Never increase the amount of washers.

Risk of temperature increase!

Risk of explosion!

Remove clamping disc of the safety diaphragm with an

open-ended wrench and remove together with safety dia-

phragm and support disc.

Remove intermediate flange by hand.

Use hex key to remove four socket head screws from

pump head and remove upper housing (housing cover

with housing cover insert and head cover).

Never remove parts by using a spiky or sharp-edged tool

(e. g. screw driver), we recommend to use a rubber mallet

or compressed air (to be blown carefully into port).

Carefully remove head cover from housing cover insert

and check valves. Note position of valves and remove.

Replace valves if necessary.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.

Replacing the diaphragm

Check diaphragm for damage and replace if necessary.

Lift diaphragm carefully sidewise.

Never use a spiky or sharp-edged tool to lift the dia-

phragm.

Use the diaphragm key to grip the diaphragm support disc

below the diaphragm.

Unscrew diaphragm support disc with diaphragm and dia-

phragm clamping disc.

Advertising