Use hex key to remove four socket head screws from

pump head and remove upper housing (housing cover

with housing cover insert and head cover).

Never remove parts by using a spiky or sharp-edged tool

(e. g. screw driver), we recommend to use a rubber mallet

or compressed air (to be blown carefully into port).

Carefully remove head cover from housing cover insert

and check valves. Note position of valves and remove.

Replace valves if necessary.

Use petroleum ether or industrial solvent to remove de-

posits. Do not inhale.

Use a face wrench to remove diaphragm support disc.

Check for washers under clamping disc. Do not mix the

washers from the different heads. Make sure that the origi-

nal number and thickness of washers is reassembled at

the individual pump head.

Never increase the amount of washers.

Risk of temperature increase!

Risk of explosion!

Remove the intermediate flange with clamping discs, di-

aphragms and diaphragm support discs from the pump

housing.

Remove the diaphragm support disc of the safety dia-

phragm and safety diaphragm.