Troubleshooting – VACUUBRAND MZ 2C EX +IK+EK User Manual

Page 31

Advertising
background image

page 31 of 75

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

Never operate a defective pump!

Read and obey the instructions for use!

Fault

Possible cause

Remedy

❑ Pump does not start or

stops immediately.

➨ Mains not plugged in, electri-

cal supply failure?

✔ Check power supply and

fuse in the building instal-

lation.

➨ Pressure in outlet pipeline too

high?

✔ Remove blockade in line,

open valve.

➨ Motor overloaded?

✔ Allow the motor to cool down,

identify cause of failure and

eliminate. Reset the motor

protection and switch on the

pump, see section above.

❑ Pump does not achieve

its ultimate vacuum or

usual pumping speed.

➨ Centring ring at small flange

connection not correctly posi-

tioned or leak in the pipeline

or vacuum system?

✔ Check pump directly - con-

nect vacuum gauge directly

at pump inlet - then check

connection, pipeline and

vacuum system if necessary.

➨ Long, narrow line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Pump has been exposed to

condensate?

✔ Allow pump to run for some

minutes with atmospheric

pressure at the inlet.

➨ Deposits have been formed

inside the pump?

✔ Clean and inspect the pump

heads.

➨ Diaphragms or valves dam-

aged?

✔ Replace diaphragms and/or

valves.

➨ Outgassing substances or

vapour generated in the

process?

✔ Check process parameters.

❑ Pump too noisy.

Attention: Switch off

pump immediately and

inspect!

➨ Loud exhaust noise?

✔ Connect hose or silencer to

pump outlet.

➨ Diaphragm crack or dia-

phragm clamping disc loose?

✔ Perform maintenance.

➨ Motor bearing defective?

✔ Contact local distributor.

➨ Deposits have been formed

inside the pump?

✔ Clean and inspect pump

heads.

➨ Diaphragms or valves defec-

tive?

✔ Replace diaphragms and/or

valves.

➨ Other than above mentioned

causes?

✔ Contact local distributor.

❑ Pump seized.

✔ Contact local distributor.

Troubleshooting