VACUUBRAND MZ 2C EX +IK+EK User Manual

Page 52

Advertising
background image

page 52 of 75

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

Inspection and replacement of the internal overpressure safety valve at the manifold

at the outlet:

(only MV 10C EX / MV 10C EX + AK + EK)

Note position of valve and remove.

Check valve for damage and replace if necessary. Make

sure that the valve is correctly seated.

Reassemble distributor and cover plate.

Elbow fitting (3x):

Slip the tubing onto the elbow fitting by turning the fitting

with an open-ended wrench (width 15 mm).

Straight fitting (2x):

Slip the tubing onto the fitting.

Tighten union nuts first by hand and then tighten

one full

turn using the open ended wrench (width 17 mm).

Use an open-ended wrench (width 17 mm) to unscrew at

the pump head the union nut of the tubing which is con-

nected to the outlet.

Turn the fitting with an open-ended wrench (width 14 mm)

to detach the tubing from the fitting (1/4 of a turn at maxi-

mum).

Do not remove the elbow fitting from the distributor cover.

Unscrew the two countersunk screws (A) at the cover plate

and remove cover plate together with distributor cover.

(A)

Note position of valve and remove.

Check valve for damage and replace if necessary. Make

sure that the valve is correctly seated.

Reassemble distributor cover and cover plate.

Slip the tubing onto the elbow fitting by turning the fitting

with an open-ended wrench (width 15 mm).

Tighten union nut first by hand and then tighten

one full

turn using the open ended wrench (width 17 mm).

Advertising
This manual is related to the following products: