Spare parts

Set of seals (diaphragms, valves, O-rings)

MZ 2C EX / MZ 2C EX + AK + EK / MZ 2C EX + IK + EK .....................................................1 x 696837

MD 4C EX / MD 4C EX + AK + EK .........................................................................................2 x 696837

MV 10C EX / MV 10C EX + AK + EK ......................................................................................4 x 696837

Sealing bond (PTFE) ..................................................................................................................637514

Valve (one piece) .........................................................................................................................637225

(internal overpressure valve in MD 4C EX, MD 4C EX + AK + EK, MV 10C EX, MV 10C EX + AK + EK)

O-Ring 28 x 2.5.............................................................................................................................635628

at the spherical ground joint of the catchpot at the inlet

Flat seal (overpressure safety valve at the outlet) ........................................................................637081

Spring (overpressure safety valve at the outlet) ...........................................................................637065

Please read section ”Replacing diaphragms and valves” com-

pletely before starting maintenance.

Partially the pictures show pumps in other versions. This doesn’t influence re-

placing diaphragms and valves of the pump.

If not using genuine spare parts the ATEX conformity becomes invalid.

Attention: Open the pump only outside of explosive atmospheres and only by

suitably trained personnel.

Always replace both diaphragms of a pump head!

Always disassemble and assemble only one pump head before opening the

next pump head. The single parts of a pump head are measured in a way so to

avoid that the diaphragm clamping disc hits the head cover. Never mix parts

of different pump heads or add or leave out washers.

We recommend to replace all diaphragms and valves of a pump at the same

time.

Attention:

The pump heads of VACUUBRAND ATEX diaphragm pumps are

assembled in two different versions. This is only visible after dis-

assembling the pump heads.

Replacing diaphragms and valves is described for both versions

separately.

Attention: The pump might be contaminated with the process chemicals that have

been pumped during operation. Ensure that the pump is decontaminated before

maintenance and take adequate precautions to protect people from the effects of

dangerous substances if contamination has occurred.

Wear appropriate safety-clothing when you come in contact with contaminated

components.